{"id":"CVE-2014-5434","published":"2019-03-26T14:59:52Z","modified":"2024-08-06T11:41:49Z","title":"Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has…","vendor":"Baxter","product":"SIGMA Spectrum Infusion System","tier":"tail","description":{"text":"Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.","provenance":"source","from":"CVE.org"},"cvss":null,"epss":{"score":0.0156,"percentile":0.75047,"provenance":"source"},"cwes":[{"id":"CWE-259","name":"Use of Hard-coded Password"}],"references":[{"url":"https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01","tags":["x_refsource_MISC"]}],"priority":{"score":15.0,"factors":[{"name":"EPSS percentile","points":15.01}]},"vendorSlug":"baxter","affectedVersions":[{"vendor":"Baxter","product":"SIGMA Spectrum Infusion System","versions":["6.05 (model 35700BAX) with wireless battery module (WBM) version 16"]}],"euvd":{"id":"EUVD-2014-5322","exploited":false,"provenance":"source"}}