{"id":"CVE-2015-3954","published":"2019-03-25T16:12:01Z","modified":"2024-08-06T06:04:01Z","title":"Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq…","vendor":"Hospira","product":"Plum A+ Infusion System","tier":"tail","description":{"text":"Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.","provenance":"source","from":"CVE.org"},"cvss":null,"epss":{"score":0.02036,"percentile":0.80956,"provenance":"source"},"cwes":[{"id":"CWE-285","name":"Improper Authorization"}],"references":[{"url":"https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01","tags":["x_refsource_MISC"]}],"priority":{"score":16.2,"factors":[{"name":"EPSS percentile","points":16.19}]},"vendorSlug":"hospira","affectedVersions":[{"vendor":"Hospira","product":"Plum A+ Infusion System","versions":["<= 13.4"]},{"vendor":"Hospira","product":"Plum A+3 Infusion System","versions":["<= 13.6"]},{"vendor":"Hospira","product":"Symbiq Infusion System","versions":["<= 3.13"]}],"euvd":{"id":"EUVD-2015-3985","exploited":false,"provenance":"source"}}