{"id":"CVE-2017-12718","published":"2018-02-15T10:00:00Z","modified":"2024-08-05T18:43:56Z","title":"A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump…","vendor":null,"product":null,"tier":"tail","description":{"text":"A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.","provenance":"source","from":"CVE.org"},"cvss":null,"epss":{"score":0.12785,"percentile":0.96374,"provenance":"source"},"cwes":[{"id":"CWE-120","name":"Buffer Copy without Checking Size of Input"}],"references":[{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A","tags":["x_refsource_MISC"]},{"url":"http://www.securityfocus.com/bid/101252","tags":["vdb-entry","x_refsource_BID"]},{"url":"http://www.securityfocus.com/bid/100665","tags":["vdb-entry","x_refsource_BID"]},{"url":"https://www.exploit-db.com/exploits/43776/","tags":["exploit","x_refsource_EXPLOIT-DB"]}],"priority":{"score":24.3,"factors":[{"name":"EPSS percentile","points":19.27},{"name":"Public exploit code","points":5.0}]},"exploitSources":{"exploitdb":{"count":1,"earliest":"2018-01-18"},"first":{"date":"2018-01-18","source":"Exploit-DB","url":"https://www.exploit-db.com/exploits/43776"},"provenance":"source"},"affectedVersions":[{"vendor":"n/a","product":"Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump","versions":["Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump"]}],"summary":{"text":"A third-party component in the Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump does not verify input buffer size before copying, resulting in a classic buffer overflow. An attacker who can get the pump to receive a crafted input can achieve remote code execution on the device, though the pump only receives such input infrequently and under certain conditions, which makes exploitation harder.","provenance":"ai","model":"claude-haiku-5-5","generatedAt":"2026-11-21T15:02:53Z"},"euvd":{"id":"EUVD-2017-4257","exploited":false,"provenance":"source"},"exploits":[{"id":43776,"url":"https://www.exploit-db.com/exploits/43776","title":"Smiths Medical Medfusion 4000 - 'DHCP' Denial of Service","date":"2018-01-18","verified":false,"provenance":"source"}],"exploitRefs":["https://www.exploit-db.com/exploits/43776/"]}