{"id":"CVE-2018-10596","published":"2018-07-02T18:00:00Z","modified":"2025-05-22T17:36:04Z","title":"Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints","vendor":"Medtronic","product":"2090 CareLink Programmer","tier":"tail","description":{"text":"Medtronic 2090 CareLink Programmer \n\nuses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.","provenance":"source","from":"CVE.org"},"cvss":{"score":7.1,"severity":"HIGH","version":"3.1","vector":"CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.0133,"percentile":0.71763,"provenance":"source"},"cwes":[{"id":"CWE-923","name":"Improper Restriction of Communication Channel to Intended Endpoints"}],"references":[{"url":"https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html","tags":[]},{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01","tags":[]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01","tags":["x_refsource_MISC","x_transferred"]}],"priority":{"score":19.3,"factors":[{"name":"EPSS percentile","points":14.35},{"name":"CVSS severity","points":4.97}]},"vendorSlug":"medtronic","affectedVersions":[{"vendor":"Medtronic","product":"2090 CareLink Programmer","versions":["All versions"]},{"vendor":"Medtronic","product":"29901 Encore Programmer","versions":["All versions"]}],"workarounds":{"items":["Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual: * Maintain good physical controls over the programmer. Having a secure physical environment prevents access to the internals of the programmer. * Only connect the programmer to managed, secure networks. * Update the software on the programmer when Medtronic updates are available. * Alternatively, disconnect the programmer from the network. Network connectivity is not required for normal programmer operation. * Offline updates are available, contact your Medtronic representative for more information. Medtronic has deployed mitigating patches to address the reported vulnerabilities.…"],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2018-2668","exploited":false,"provenance":"source"}}