{"id":"CVE-2018-10626","published":"2018-08-10T18:00:00Z","modified":"2026-05-19T14:32:21Z","title":"Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity","vendor":"Medtronic","product":"24950 MyCareLink Monitor","tier":"tail","description":{"text":"Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An \nattacker who obtains per-product credentials from the monitor and paired\n implantable cardiac device information can potentially upload invalid \ndata to the Medtronic CareLink network.","provenance":"source","from":"CVE.org"},"cvss":{"score":4.4,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:N","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00361,"percentile":0.30731,"provenance":"source"},"cwes":[{"id":"CWE-345","name":"Insufficient Verification of Data Authenticity"}],"references":[{"url":"https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-18.html","tags":[]},{"url":"http://www.securityfocus.com/bid/105042","tags":["vdb-entry"]},{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-219-01","tags":[]},{"url":"https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2018/icsma-18-219-01.json","tags":[]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01","tags":["x_refsource_MISC","x_transferred"]}],"priority":{"score":9.2,"factors":[{"name":"EPSS percentile","points":6.15},{"name":"CVSS severity","points":3.08}]},"vendorSlug":"medtronic","affectedVersions":[{"vendor":"Medtronic","product":"24950 MyCareLink Monitor","versions":["All versions"]},{"vendor":"Medtronic","product":"24952 MyCareLink Monitor","versions":["All versions"]}],"workarounds":{"items":["Medtronic has made server-side updates to address this insufficient verification vulnerability. Medtronic is implementing additional server-side mitigations to enhance data integrity and authenticity.","Medtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should: * Maintain good physical control over the home monitor. * Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system.","Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security","Users should follow CISA's guidance in the following areas: * Securing the Internet of Things https://www.cisa.gov/news-events/news/securing-internet-things-iot * Home Network Security https://www.cisa.gov/news-events/news/home-network-security"],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2018-2698","exploited":false,"provenance":"source"}}