{"id":"CVE-2018-10634","published":"2018-08-13T22:00:00Z","modified":"2025-05-22T16:28:03Z","title":"Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Cleartext Transmission of Sensitive Information","vendor":"Medtronic","product":"MMT- 508 - MiniMed pump","tier":"tail","description":{"text":"Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.","provenance":"source","from":"CVE.org"},"cvss":{"score":4.8,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:A/AC:H/PR:N/UI:R/S:U/C:H/I:N/A:N","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00474,"percentile":0.41917,"provenance":"source"},"cwes":[{"id":"CWE-319","name":"Cleartext Transmission of Sensitive Information"}],"references":[{"url":"https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html","tags":[]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02","tags":[]},{"url":"http://www.securityfocus.com/bid/105044","tags":["vdb-entry"]}],"priority":{"score":11.7,"factors":[{"name":"EPSS percentile","points":8.38},{"name":"CVSS severity","points":3.36}]},"vendorSlug":"medtronic","affectedVersions":[{"vendor":"Medtronic","product":"MMT- 508 - MiniMed pump","versions":["All versions"]},{"vendor":"Medtronic","product":"MMT – 511 pump Paradigm","versions":["All versions"]},{"vendor":"Medtronic","product":"MMT – 512 / MMT – 712 Paradigm x12","versions":["All versions"]},{"vendor":"Medtronic","product":"MMT – 515 / MMT – 715 Paradigm x15","versions":["All versions"]},{"vendor":"Medtronic","product":"MMT – 522 / MMT – 722 Paradigm REAL-TIME","versions":["All versions"]},{"vendor":"Medtronic","product":"MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME","versions":["All versions"]},{"vendor":"Medtronic","product":"MMT – 523 / MMT – 723 Paradigm Revel","versions":["All versions"]},{"vendor":"Medtronic","product":"MMT – 523(K) / MMT – 723(K) Paradigm","versions":["All versions"]},{"vendor":"Medtronic","product":"MMT – 554 / MMT – 754 MiniMed Veo","versions":["All versions"]},{"vendor":"Medtronic","product":"MMT – 551 / MMT – 751 MiniMed 530G","versions":["All versions"]}],"workarounds":{"items":["The remote option is turned off in the pump by default. Medtronic is directing all users to stop using their remote controllers, disable the remote option on their insulin pump, and to return the remote controllers to Medtronic. Medtronic has released additional patient focused information https://www.medtronic.com/security . Additionally, Medtronic will be sending a letter to patients who may still be actively using the remotes in order to inform patients about these security risks, and request patients stop using the remote and return them to Medtronic."],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2018-2706","exploited":false,"provenance":"source"}}