{"id":"CVE-2018-18984","published":"2018-12-14T15:00:00Z","modified":"2025-05-22T16:40:53Z","title":"Medtronic 9790, 2090 CareLink, and 29901 Encore Programmers Missing Encryption of Sensitive Data","vendor":"Medtronic","product":"CareLink 9790 Programmer","tier":"tail","description":{"text":"Medtronic CareLink and Encore Programmers\n\n do not encrypt or do not sufficiently encrypt sensitive \nPII and PHI information while at rest .","provenance":"source","from":"CVE.org"},"cvss":{"score":4.6,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00315,"percentile":0.25118,"provenance":"source"},"cwes":[{"id":"CWE-311","name":"Missing Encryption of Sensitive Data"}],"references":[{"url":"https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-9790-2090-29901.html","tags":[]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-347-01","tags":[]},{"url":"http://www.securityfocus.com/bid/106215","tags":["vdb-entry"]}],"priority":{"score":8.2,"factors":[{"name":"EPSS percentile","points":5.02},{"name":"CVSS severity","points":3.22}]},"vendorSlug":"medtronic","affectedVersions":[{"vendor":"Medtronic","product":"CareLink 9790 Programmer","versions":["All versions"]},{"vendor":"Medtronic","product":"CareLink 2090 Programmer","versions":["All versions"]},{"vendor":"Medtronic","product":"29901 Encore Programmer","versions":["All versions"]}],"workarounds":{"items":["The CareLink 9790 Programmer has been placed into end-of-life status and is no longer supported by Medtronic. Medtronic recommends users no longer use the 9790 for any purpose. The CareLink 2090 and 29901 Encore programmers store PHI and PII as part of their normal operating procedure. Medtronic recommends that when devices are storing PHI/PII it should be retained on these programmers for the least amount of time necessary, and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy. Please contact a Medtronic representative for proper disposal and PHI/PII retention setting instructions. All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer. Medtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, hospitals and clinicians should: * Maintain strict physical control of the programmer. * Use only legitimately obtained programmers and not ones provided by any third party. Proper disposal of…"],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2018-10688","exploited":false,"provenance":"source"}}