{"id":"CVE-2018-5446","published":"2018-05-04T18:00:00Z","modified":"2025-05-22T17:40:34Z","title":"Medtronic 2090 Carelink Programmer Storing Passwords in a Recoverable Format","vendor":"Medtronic","product":"2090 CareLink Programmer","tier":"tail","description":{"text":"Medtronic 2090 CareLink Programmer \n\nuses a per-product username and password that is stored in a recoverable format.","provenance":"source","from":"CVE.org"},"cvss":{"score":4.9,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00385,"percentile":0.33473,"provenance":"source"},"cwes":[{"id":"CWE-257","name":"Storing Passwords in a Recoverable Format"}],"references":[{"url":"https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html","tags":[]},{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01","tags":[]},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01","tags":["x_refsource_MISC","x_transferred"]}],"priority":{"score":10.1,"factors":[{"name":"EPSS percentile","points":6.69},{"name":"CVSS severity","points":3.43}]},"vendorSlug":"medtronic","affectedVersions":[{"vendor":"Medtronic","product":"2090 CareLink Programmer","versions":["All versions"]},{"vendor":"Medtronic","product":"29901 Encore Programmer","versions":["All versions"]}],"workarounds":{"items":["Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual: * Maintain good physical controls over the programmer. Having a secure physical environment prevents access to the internals of the programmer. * Only connect the programmer to managed, secure networks. * Update the software on the programmer when Medtronic updates are available. * Alternatively, disconnect the programmer from the network. Network connectivity is not required for normal programmer operation. * Offline updates are available, contact your Medtronic representative for more information. Medtronic has deployed mitigating patches to address the reported vulnerabilities.…"],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2018-17216","exploited":false,"provenance":"source"}}