{"id":"CVE-2019-25720","published":"2026-06-03T16:56:29Z","modified":"2026-06-03T17:44:47Z","title":"Dräger SC Monitoring Devices DoS via Malformed Network Packet","vendor":"Dräger","product":"SC 6002XL","tier":"tail","description":{"text":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.","provenance":"source","from":"CVE.org"},"cvss":{"score":6.5,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00199,"percentile":0.09967,"provenance":"source"},"cwes":[{"id":"CWE-1286","name":"Improper Validation of Syntactic Correctness of Input"}],"references":[{"url":"https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf","tags":["vendor-advisory"]},{"url":"https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet","tags":["third-party-advisory"]}],"priority":{"score":6.5,"factors":[{"name":"EPSS percentile","points":1.99},{"name":"CVSS severity","points":4.55}]},"vendorSlug":"drager","affectedVersions":[{"vendor":"Dräger","product":"SC 6002XL","versions":["SC 6002XL"]},{"vendor":"Dräger","product":"SC6802XL","versions":["SC6802XL"]},{"vendor":"Dräger","product":"SC 7000","versions":["SC 7000"]},{"vendor":"Dräger","product":"SC8000","versions":["SC8000"]},{"vendor":"Dräger","product":"SC90000 XL","versions":["SC90000 XL"]}],"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"partial","timestamp":"2026-06-03T17:44:08.218138Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2019-20161","exploited":false,"provenance":"source"}}