{"id":"CVE-2019-25721","published":"2026-06-02T17:29:08Z","modified":"2026-06-03T16:06:32Z","title":"Dräger Infinity M300 VG2.3.1 Network-Based Denial of Service","vendor":"Dräger","product":"Infinity M300","tier":"tail","description":{"text":"Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.","provenance":"source","from":"CVE.org"},"cvss":{"score":6.5,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00187,"percentile":0.08534,"provenance":"source"},"cwes":[{"id":"CWE-400","name":"Uncontrolled Resource Consumption"}],"references":[{"url":"https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf","tags":["vendor-advisory"]},{"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service","tags":["third-party-advisory"]}],"priority":{"score":6.3,"factors":[{"name":"EPSS percentile","points":1.71},{"name":"CVSS severity","points":4.55}]},"vendorSlug":"drager","affectedVersions":[{"vendor":"Dräger","product":"Infinity M300","versions":["<= VG2.3.1"]}],"fixedVersions":[{"vendor":"Dräger","product":"Infinity M300","versions":["VG2.3.2"]}],"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"partial","timestamp":"2026-06-03T12:52:13.424163Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2019-20157","exploited":false,"provenance":"source"}}