{"id":"CVE-2019-25723","published":"2026-06-02T19:00:19Z","modified":"2026-06-03T18:01:04Z","title":"Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface","vendor":"Dräger","product":"Perseus A500","tier":"tail","description":{"text":"Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.","provenance":"source","from":"CVE.org"},"cvss":{"score":4,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00236,"percentile":0.14939,"provenance":"source"},"cwes":[{"id":"CWE-1286","name":"Improper Validation of Syntactic Correctness of Input"}],"references":[{"url":"https://static.draeger.com/security/download/2020-06-PSA-20-169-01-Security-Advisory-Perseus-A500-SW20n.pdf","tags":["vendor-advisory"]},{"url":"https://www.vulncheck.com/advisories/dr-ger-perseus-a500-dos-via-medibus-interface","tags":["third-party-advisory"]}],"priority":{"score":5.8,"factors":[{"name":"EPSS percentile","points":2.99},{"name":"CVSS severity","points":2.8}]},"vendorSlug":"drager","affectedVersions":[{"vendor":"Dräger","product":"Perseus A500","versions":["2.00","2.01","2.02"]}],"fixedVersions":[{"vendor":"Dräger","product":"Perseus A500","versions":["2.03"]}],"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"partial","timestamp":"2026-06-03T13:56:22.966552Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2019-20159","exploited":false,"provenance":"source"}}