{"id":"CVE-2020-12035","published":"2020-06-29T13:49:50Z","modified":"2024-08-04T11:48:57Z","title":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service…","vendor":null,"product":null,"tier":"tail","description":{"text":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.","provenance":"source","from":"CVE.org"},"cvss":null,"epss":{"score":0.00247,"percentile":0.09633,"provenance":"source"},"cwes":[{"id":"CWE-287","name":"Improper Authentication"}],"references":[{"url":"https://www.us-cert.gov/ics/advisories/icsma-20-170-01","tags":["x_refsource_MISC"]}],"priority":{"score":1.9,"factors":[{"name":"EPSS percentile","points":1.93}]},"affectedVersions":[{"vendor":"n/a","product":"Baxter PrismaFlex and PrisMax","versions":["PrismaFlex all versions, PrisMax all versions prior to 3.x"]}],"euvd":{"id":"EUVD-2020-4351","exploited":false,"provenance":"source"}}