{"id":"CVE-2020-16216","published":"2020-09-11T13:06:55Z","modified":"2024-08-04T13:37:54Z","title":"Philips Patient Monitoring Devices Improper Input Validation","vendor":"Philips","product":"IntelliVue patient monitors","tier":"tail","description":{"text":"In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, \nMX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, \nthe product receives input or data but does not validate or incorrectly \nvalidates that the input has the properties required to process the data\n safely and correctly, which can induce a denial-of-service condition \nthrough a system restart.","provenance":"source","from":"CVE.org"},"cvss":null,"epss":{"score":0.00698,"percentile":0.54022,"provenance":"source"},"cwes":[{"id":"CWE-20","name":"Improper Input Validation"}],"references":[{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01","tags":["x_refsource_MISC"]},{"url":"https://www.philips.com/productsecurity","tags":[]}],"priority":{"score":10.8,"factors":[{"name":"EPSS percentile","points":10.8}]},"vendorSlug":"philips","affectedVersions":[{"vendor":"Philips","product":"IntelliVue patient monitors","versions":["MX100","MX400-550","MX600","MX700","MX750","MX800","MX850","MP2-MP90"]},{"vendor":"Philips","product":"IntelliVue","versions":["X2","X3","<= N"]}],"solutions":{"items":["Philips released the following versions to remediate reported vulnerabilities: * IntelliVue Patient Monitors Versions N.00 and N.01 * IntelliVue Patient Monitors Version M.04: Contact a Philips service support team https://www.usa.philips.com/healthcare/solutions/customer-service-solutions for an upgrade path * Certificate revocation implementation of the IntelliVue Patient Monitors will be completed in Q3 of 2024."],"from":"CVE.org CNA","provenance":"source"},"workarounds":{"items":["As a mitigation to these vulnerabilities, Philips recommends the following: * The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses. Refer to the Philips Patient Monitoring System Security for Clinical Networks guide for additional information on InCenter https://incenter.medical.philips.com/ . * By default, the simple certificate enrollment protocol (SCEP) service is not running. When needed, the service is configured to run based on the duration or the number of certificates to be assigned. One certificate is default, but if a certificate is not issued, the service will continue to run. Limit exposure by ensuring the SCEP service is not running unless it is actively being used to enroll new devices. * When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits. * Implement physical security controls to prevent unauthorized login attempts on the PIC iX…"],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2020-8182","exploited":false,"provenance":"source"}}