{"id":"CVE-2020-16218","published":"2020-09-11T12:50:08Z","modified":"2024-08-04T13:37:54Z","title":"Philips Patient Monitoring Devices Cross-site Scripting","vendor":"Philips","product":"Patient Information Center iX (PICiX)","tier":"tail","description":{"text":"In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the \nsoftware does not neutralize or incorrectly neutralizes \nuser-controllable input before it is placed in output that is then used \nas a webpage and served to other users. Successful exploitation could \nlead to unauthorized access to patient data via a read-only web \napplication.","provenance":"source","from":"CVE.org"},"cvss":null,"epss":{"score":0.00653,"percentile":0.52226,"provenance":"source"},"cwes":[{"id":"CWE-79","name":"Improper Neutralization of Input During Web Page Generation"}],"references":[{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01","tags":["x_refsource_MISC"]},{"url":"https://www.philips.com/productsecurity","tags":[]}],"priority":{"score":10.4,"factors":[{"name":"EPSS percentile","points":10.45}]},"vendorSlug":"philips","affectedVersions":[{"vendor":"Philips","product":"Patient Information Center iX (PICiX)","versions":["B.02","C.02","C.03"]}],"solutions":{"items":["Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * Certificate revocation within the system was implemented for PIC iX."],"from":"CVE.org CNA","provenance":"source"},"workarounds":{"items":["As a mitigation to these vulnerabilities, Philips recommends the following: * The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses. Refer to the Philips Patient Monitoring System Security for Clinical Networks guide for additional information on InCenter https://incenter.medical.philips.com/ . * By default, the simple certificate enrollment protocol (SCEP) service is not running. When needed, the service is configured to run based on the duration or the number of certificates to be assigned. One certificate is default, but if a certificate is not issued, the service will continue to run. Limit exposure by ensuring the SCEP service is not running unless it is actively being used to enroll new devices. * When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits. * Implement physical security controls to prevent unauthorized login attempts on the PIC iX…"],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2020-8184","exploited":false,"provenance":"source"}}