{"id":"CVE-2020-16228","published":"2020-09-11T12:58:13Z","modified":"2024-08-04T13:37:54Z","title":"Philips Patient Monitoring Devices Improper Check for Certificate Revocation","vendor":"Philips","product":"Patient Information Center iX (PICiX)","tier":"tail","description":{"text":"In Patient Information Center iX (PICiX) Versions C.02 and C.03, \nPerformanceBridge Focal Point Version A.01, IntelliVue patient monitors \nMX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and \nprior, the software does not check or incorrectly checks the revocation \nstatus of a certificate, which may cause it to use a compromised \ncertificate.","provenance":"source","from":"CVE.org"},"cvss":null,"epss":{"score":0.00368,"percentile":0.31488,"provenance":"source"},"cwes":[{"id":"CWE-299","name":"Improper Check for Certificate Revocation"}],"references":[{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01","tags":["x_refsource_MISC"]},{"url":"https://www.philips.com/productsecurity","tags":[]}],"priority":{"score":6.3,"factors":[{"name":"EPSS percentile","points":6.3}]},"vendorSlug":"philips","affectedVersions":[{"vendor":"Philips","product":"Patient Information Center iX (PICiX)","versions":["C.02","C.03"]},{"vendor":"Philips","product":"PerformanceBridge Focal Point","versions":["A.01"]},{"vendor":"Philips","product":"IntelliVue patient monitors","versions":["MX100","MX400-MX550","MX750","MX850"]},{"vendor":"Philips","product":"IntelliVue X3","versions":["<= N"]}],"solutions":{"items":["Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * PerformanceBridge Focal Point * IntelliVue Patient Monitors Versions N.00 and N.01 * IntelliVue Patient Monitors Version M.04: Contact a Philips service support team https://www.usa.philips.com/healthcare/solutions/customer-service-solutions for an upgrade path * Certificate revocation within the system was implemented for PIC iX and Performance Bridge FocalPoint in 2023. The implementation of the IntelliVue Patient Monitors will be completed in Q3 of 2024."],"from":"CVE.org CNA","provenance":"source"},"workarounds":{"items":["As a mitigation to these vulnerabilities, Philips recommends the following: * The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses. Refer to the Philips Patient Monitoring System Security for Clinical Networks guide for additional information on InCenter https://incenter.medical.philips.com/ . * By default, the simple certificate enrollment protocol (SCEP) service is not running. When needed, the service is configured to run based on the duration or the number of certificates to be assigned. One certificate is default, but if a certificate is not issued, the service will continue to run. Limit exposure by ensuring the SCEP service is not running unless it is actively being used to enroll new devices. * When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits. * Implement physical security controls to prevent unauthorized login attempts on the PIC iX…"],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2020-8194","exploited":false,"provenance":"source"}}