{"id":"CVE-2021-33882","published":"2021-08-25T11:19:46Z","modified":"2024-08-04T00:05:51Z","title":"A Missing Authentication for Critical Function vulnerability in B","vendor":null,"product":null,"tier":"tail","description":{"text":"A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.","provenance":"source","from":"CVE.org"},"cvss":{"score":6.8,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AC:H/AV:N/A:N/C:N/I:H/PR:N/S:C/UI:N","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.011,"percentile":0.65557,"provenance":"source"},"cwes":[],"references":[{"url":"https://www.bbraunusa.com/en.htm","tags":["x_refsource_MISC"]},{"url":"https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/","tags":["x_refsource_MISC"]}],"priority":{"score":17.9,"factors":[{"name":"EPSS percentile","points":13.11},{"name":"CVSS severity","points":4.76}]},"euvd":{"id":"EUVD-2021-20555","exploited":false,"provenance":"source"}}