{"id":"CVE-2021-33885","published":"2021-08-25T11:38:20Z","modified":"2024-08-04T00:05:52Z","title":"An Insufficient Verification of Data Authenticity vulnerability in B","vendor":null,"product":null,"tier":"tail","description":{"text":"An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.","provenance":"source","from":"CVE.org"},"cvss":{"score":10,"severity":"CRITICAL","version":"3.1","vector":"CVSS:3.1/AC:L/AV:N/A:N/C:H/I:H/PR:N/S:C/UI:N","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.0558,"percentile":0.93068,"provenance":"source"},"cwes":[],"references":[{"url":"https://www.bbraunusa.com/en.htm","tags":["x_refsource_MISC"]},{"url":"https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/","tags":["x_refsource_MISC"]}],"priority":{"score":25.6,"factors":[{"name":"EPSS percentile","points":18.61},{"name":"CVSS severity","points":7.0}]},"summary":{"text":"The flaw is an insufficient verification of data authenticity in B. Braun SpaceCom2 (versions prior to 012U000062), where critical data sets lack cryptographic signatures. A remote, unauthenticated attacker with network access can send malicious data that the device uses in place of the correct data, gaining full system command access and execution, with no user interaction or authentication required.","provenance":"ai","model":"claude-haiku-5-5","generatedAt":"2026-11-20T06:31:52Z"},"euvd":{"id":"EUVD-2021-20558","exploited":false,"provenance":"source"}}