{"id":"CVE-2021-38392","published":"2021-10-04T17:35:13Z","modified":"2024-09-17T03:33:08Z","title":"Improper Access Control for Boston Scientific Zoom Latitude","vendor":"Boston Scientific","product":"ZOOM LATITUDE","tier":"tail","description":{"text":"A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.","provenance":"source","from":"CVE.org"},"cvss":{"score":6.5,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00259,"percentile":0.10817,"provenance":"source"},"cwes":[{"id":"CWE-284","name":"Improper Access Control"}],"references":[{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01","tags":["x_refsource_MISC"]}],"priority":{"score":6.7,"factors":[{"name":"EPSS percentile","points":2.16},{"name":"CVSS severity","points":4.55}]},"vendorSlug":"boston-scientific","affectedVersions":[{"vendor":"Boston Scientific","product":"ZOOM LATITUDE","versions":["Model 3120"]}],"workarounds":{"items":["Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2021-24844","exploited":false,"provenance":"source"}}