{"id":"CVE-2021-38394","published":"2021-10-04T17:34:58Z","modified":"2024-09-16T18:59:19Z","title":"Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging Techniques for Boston…","vendor":"Boston Scientific","product":"ZOOM LATITUDE","tier":"tail","description":{"text":"An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.","provenance":"source","from":"CVE.org"},"cvss":{"score":6.2,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:L","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00234,"percentile":0.08667,"provenance":"source"},"cwes":[{"id":"CWE-1278","name":"Missing Protection Against Hardware Reverse Engineering Using Integrated Circuit (IC) Imaging Techniques"}],"references":[{"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01","tags":["x_refsource_MISC"]}],"priority":{"score":6.1,"factors":[{"name":"EPSS percentile","points":1.73},{"name":"CVSS severity","points":4.34}]},"vendorSlug":"boston-scientific","affectedVersions":[{"vendor":"Boston Scientific","product":"ZOOM LATITUDE","versions":["Model 3120"]}],"workarounds":{"items":["Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2021-24846","exploited":false,"provenance":"source"}}