{"id":"CVE-2021-43935","published":"2021-12-15T18:05:16Z","modified":"2024-09-16T23:11:47Z","title":"ICSMA-21-343-01 Hillrom Welch Allyn Cardio Products","vendor":"Hillrom","product":"Welch Allyn Q-Stress Cardiac Stress Testing System","tier":"tail","description":{"text":"The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. This vulnerability allows the application to accept manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges.","provenance":"source","from":"CVE.org"},"cvss":{"score":8.1,"severity":"HIGH","version":"3.1","vector":"CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.01145,"percentile":0.67525,"provenance":"source"},"cwes":[{"id":"CWE-288","name":"Authentication Bypass Using an Alternate Path or Channel"}],"references":[{"url":"https://www.cisa.gov/uscert/ics/advisories/icsma-21-343-01","tags":["x_refsource_MISC"]}],"priority":{"score":19.2,"factors":[{"name":"EPSS percentile","points":13.51},{"name":"CVSS severity","points":5.67}]},"vendorSlug":"hillrom","affectedVersions":[{"vendor":"Hillrom","product":"Welch Allyn Q-Stress Cardiac Stress Testing System","versions":[">= 6.0.0, <= 6.3.1"]},{"vendor":"Hillrom","product":"Welch Allyn X-Scribe Cardiac Stress Testing System","versions":[">= 5.01, <= 6.3.1"]},{"vendor":"Hillrom","product":"Welch Allyn Diagnostic Cardiology Suite","versions":["2.1.0"]},{"vendor":"Hillrom","product":"Welch Allyn Vision Express","versions":[">= 6.1.0, <= 6.4.0"]},{"vendor":"Hillrom","product":"Welch Allyn H-Scribe Holter Analysis System","versions":[">= 5.01, <= 6.4.0"]},{"vendor":"Hillrom","product":"Welch Allyn R-Scribe Resting ECG System","versions":[">= 5.01, <= 7.0.0"]},{"vendor":"Hillrom","product":"Welch Allyn Connex Cardio","versions":[">= 1.0.0, <= 1.1.1"]}],"solutions":{"items":["Hillrom recommends users upgrade to the latest product versions when updated products are available. Information on how to update these products to their new versions can be found on the Hillrom disclosure page."],"from":"CVE.org CNA","provenance":"source"},"workarounds":{"items":["-Disable the SSO feature in the respective Modality Manager Configuration settings. Please refer to the instructions for use (IFU) and/or service manual for instructions on how to disable SSO. -Apply proper network and physical security controls. -Apply authentication for server access."],"from":"CVE.org CNA","provenance":"source"},"euvd":{"id":"EUVD-2021-30797","exploited":false,"provenance":"source"}}