{"id":"CVE-2022-26388","published":"2025-02-07T17:06:30Z","modified":"2025-02-07T18:50:58Z","title":"Use of Hard-Coded Password Vulnerability in ELI Electrocardiograph Devices","vendor":"Welch Allyn","product":"ELI 380 Resting Electrocardiograph","tier":"tail","description":{"text":"A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph: \n\nVersions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:\n\nVersions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph:\n\nVersions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: \n\nVersions 2.2.0 and prior.","provenance":"source","from":"CVE.org"},"cvss":{"score":6.4,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:L","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00274,"percentile":0.20209,"provenance":"source"},"cwes":[{"id":"CWE-259","name":"Use of Hard-coded Password"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-22-167-01","tags":[]},{"url":"https://hillrom.com/en/responsible-disclosures/","tags":[]}],"priority":{"score":8.5,"factors":[{"name":"EPSS percentile","points":4.04},{"name":"CVSS severity","points":4.48}]},"vendorSlug":"welch-allyn","affectedVersions":[{"vendor":"Welch Allyn","product":"ELI 380 Resting Electrocardiograph","versions":["<= 2.6.0"]},{"vendor":"Welch Allyn","product":"ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph","versions":["<= 2.3.1"]},{"vendor":"Welch Allyn","product":"ELI 250c/BUR 250c Resting Electrocardiograph","versions":["<= 2.1.2"]},{"vendor":"Welch Allyn","product":"ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph","versions":["<= 2.2.0"]}],"solutions":{"items":["Hillrom has released software updates for all impacted devices to address these vulnerabilities. New product versions that mitigate these vulnerabilities are available as follows: * Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023 * Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022 * Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023 Hillrom recommends users upgrade to the latest product versions. Information on how to update these products can be found on the Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ ."],"from":"CVE.org CNA","provenance":"source"},"workarounds":{"items":["Hillrom recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured for ELI Link and Cardiograph. * Where possible, use a firewall to prevent communication on Port 21 FTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet service."],"from":"CVE.org CNA","provenance":"source"},"capec":[{"id":"CAPEC-114","name":"Authentication Abuse"}],"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total","timestamp":"2025-02-07T18:50:50.198188Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2022-30947","exploited":false,"provenance":"source"}}