{"id":"CVE-2024-12248","published":"2025-01-30T18:17:22Z","modified":"2025-02-12T16:53:16Z","title":"Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor","vendor":"Contec Health","product":"CMS8000 Patient Monitor","tier":"core","eol":null,"affectedVersions":[{"vendor":"Contec Health","product":"CMS8000 Patient Monitor","versions":["Firmware version smart3250-2.6.27-wlan2.1.7.cramfs","Firmware version CMS7.820.075.08/0.74(0.75)","Firmware version CMS7.820.120.01/0.93(0.95)"]}],"solutions":{"items":["Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."],"from":"CVE.org CNA","provenance":"source"},"ssvc":{"exploitation":"none","automatable":"yes","technicalImpact":"total","timestamp":"2025-01-31T19:13:06.864223Z","from":"CISA ADP","provenance":"source"},"description":{"text":"Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.","provenance":"source","from":"CVE.org"},"summary":{"text":"The out-of-bounds write flaw lives in the Contec Health CMS8000 Patient Monitor, a patient monitoring device. An attacker who can reach the device over the network can send specially formatted UDP requests that write arbitrary data, potentially achieving remote code execution, with no authentication or user interaction required and no special configuration mentioned.","provenance":"ai","model":"claude-haiku-5-5","generatedAt":"2026-11-01T13:55:24Z"},"cvss":{"score":9.8,"severity":"CRITICAL","version":"3.1","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.01479,"percentile":0.74148,"date":"2026-11-25","provenance":"source"},"kev":null,"vulncheckKev":{"dateAdded":"2026-01-20","exploitedFirstSeen":"2026-01-20","ransomware":false,"evidence":[{"url":"https://www.loginsoft.com/reports/annually/vulnerability-intelligence-report-2025","date":"2026-01-20"}],"evidenceTotal":1,"provenance":"source"},"exploited":true,"otx":{"pulseCount":1,"pulses":[{"name":"Healthcare Sector Threat Intel - FDA 510(k) Medical Device Security (DugganUSA)","url":"https://otx.alienvault.com/pulse/692c9d81b4e8f82f6d98e1c6","created":"2025-11-30"}],"provenance":"machine"},"cwes":[{"id":"CWE-787","name":"Out-of-bounds Write"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01","tags":[]},{"url":"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication","tags":[]}],"techniques":[],"reports":[],"affected":[],"advisories":[],"euvd":{"id":"EUVD-2024-50712","url":"https://euvd.enisa.europa.eu/vulnerability/EUVD-2024-50712","exploited":false,"exploitedSince":null,"provenance":"source"},"exploits":[],"exploitSources":null,"exploitRefs":[],"flow":[],"timeline":[{"date":"2025-01-30","kind":"published","label":"Published (CVE.org)","url":"https://www.cve.org/CVERecord?id=CVE-2024-12248","provenance":"source"},{"date":"2026-01-20","kind":"exploited-reported","label":"Exploitation first reported (VulnCheck KEV evidence)","url":"https://www.loginsoft.com/reports/annually/vulnerability-intelligence-report-2025","provenance":"source"}],"speed":{"daysToPublicExploit":null,"firstExploitFrom":null,"daysToKev":null},"priority":{"score":62.4,"factors":[{"name":"In a KEV catalog","points":35.0,"detail":"listed in VulnCheck KEV"},{"name":"Recent KEV listing","points":5.0,"detail":"added 309 days ago"},{"name":"Known ransomware use","points":0.0,"detail":"not known"},{"name":"EPSS percentile","points":14.83,"detail":"percentile 0.7415"},{"name":"Named in threat reports","points":0.0,"detail":"0 reports"},{"name":"Public exploit code","points":0.0,"detail":"none found"},{"name":"Mentioned in OTX pulses","points":0.68,"detail":"1 pulse"},{"name":"CVSS severity","points":6.86,"detail":"base score 9.8"},{"name":"End-of-life product","points":0.0,"detail":"supported or unknown"}]}}