{"id":"CVE-2024-22100","published":"2024-03-01T00:37:47Z","modified":"2024-08-12T20:20:19Z","title":"MicroDicom DICOM Heap-based Buffer Overflow","vendor":"MicroDicom","product":"DICOM Viewer","tier":"tail","description":{"text":"\n\n\n\n\nMicroDicom DICOM Viewer versions 2023.3 (Build 9342) and prior are affected by a heap-based buffer overflow vulnerability, which could allow an attacker to execute arbitrary code on affected installations of DICOM Viewer. A user must open a malicious DCM file in order to exploit the vulnerability.\n\n\n\n\n\n","provenance":"source","from":"CVE.org"},"cvss":{"score":7.8,"severity":"HIGH","version":"3.1","vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00354,"percentile":0.21543,"provenance":"source"},"cwes":[{"id":"CWE-122","name":"Heap-based Buffer Overflow"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-060-01","tags":[]}],"priority":{"score":9.8,"factors":[{"name":"EPSS percentile","points":4.31},{"name":"CVSS severity","points":5.46}]},"vendorSlug":"microdicom","affectedVersions":[{"vendor":"MicroDicom","product":"DICOM Viewer","versions":["<=  2023.3 (Build 9342)"]}],"solutions":{"items":["MicroDicom has provided a fix and recommends users upgrade to 2024.1 https://www.microdicom.com/ . For additional assistance, contact MicroDicom https://www.microdicom.com/contacts.html  directly"],"from":"CVE.org CNA","provenance":"source"},"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total","timestamp":"2024-03-04T20:39:49.457575Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2024-19696","exploited":false,"provenance":"source"}}