{"id":"CVE-2024-48970","published":"2024-11-14T21:31:14Z","modified":"2024-11-18T15:23:48Z","title":"Life2000 Ventilator microcontroller lacks memory protection","vendor":"Baxter","product":"Life2000 Ventilation System","tier":"tail","description":{"text":"The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.","provenance":"source","from":"CVE.org"},"cvss":{"score":9.3,"severity":"CRITICAL","version":"3.1","vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00271,"percentile":0.12163,"provenance":"source"},"cwes":[{"id":"CWE-1191","name":"On-Chip Debug and Test Interface With Improper Access Control"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01","tags":[]}],"priority":{"score":8.9,"factors":[{"name":"EPSS percentile","points":2.43},{"name":"CVSS severity","points":6.51}]},"vendorSlug":"baxter","affectedVersions":[{"vendor":"Baxter","product":"Life2000 Ventilation System","versions":["06.08.00.00 and prior"]}],"capec":[{"id":"CAPEC-117","name":"Interception"},{"id":"CAPEC-458","name":"Flash Memory Attacks"}],"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total","timestamp":"2024-11-18T15:22:31.746766Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2024-43155","exploited":false,"provenance":"source"}}