{"id":"CVE-2024-48974","published":"2024-11-14T21:27:10Z","modified":"2024-11-18T15:32:38Z","title":"Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates","vendor":"Baxter","product":"Life2000 Ventilation System","tier":"tail","description":{"text":"The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.","provenance":"source","from":"CVE.org"},"cvss":{"score":9.3,"severity":"CRITICAL","version":"3.1","vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00154,"percentile":0.04592,"provenance":"source"},"cwes":[{"id":"CWE-494","name":"Download of Code Without Integrity Check"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01","tags":[]}],"priority":{"score":7.4,"factors":[{"name":"EPSS percentile","points":0.92},{"name":"CVSS severity","points":6.51}]},"vendorSlug":"baxter","affectedVersions":[{"vendor":"Baxter","product":"Life2000 Ventilation System","versions":["06.08.00.00 and prior"]}],"capec":[{"id":"CAPEC-186","name":"Malicious Software Update"},{"id":"CAPEC-117","name":"Interception"}],"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total","timestamp":"2024-11-18T15:31:47.610421Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2024-43158","exploited":false,"provenance":"source"}}