{"id":"CVE-2025-0626","published":"2025-01-30T18:17:29Z","modified":"2025-03-01T17:12:24Z","title":"Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor","vendor":"Contec Health","product":"CMS8000 Patient Monitor","tier":"core","eol":null,"affectedVersions":[{"vendor":"Contec Health","product":"CMS8000 Patient Monitor","versions":["All versions"]}],"workarounds":{"items":["Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."],"from":"CVE.org CNA","provenance":"source"},"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total","timestamp":"2025-01-30T19:08:00.729447Z","from":"CISA ADP","provenance":"source"},"description":{"text":"The \"monitor\" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.","provenance":"source","from":"CVE.org"},"summary":null,"cvss":{"score":7.5,"severity":"HIGH","version":"3.1","vector":"CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.01141,"percentile":0.6743,"date":"2026-11-20","provenance":"source"},"kev":null,"vulncheckKev":{"dateAdded":"2026-01-20","exploitedFirstSeen":"2026-01-20","ransomware":false,"evidence":[{"url":"https://www.loginsoft.com/reports/annually/vulnerability-intelligence-report-2025","date":"2026-01-20"}],"evidenceTotal":1,"provenance":"source"},"exploited":true,"otx":{"pulseCount":2,"pulses":[{"name":"Contec CMS8000 Contains a Backdoor","url":"https://otx.alienvault.com/pulse/67a190facf12f98043d32b8a","created":"2025-02-04"},{"name":"Contec CMS8000 Contains a Backdoor","url":"https://otx.alienvault.com/pulse/67a06a13cefb52d72853cb73","created":"2025-02-03"}],"provenance":"machine"},"cwes":[{"id":"CWE-912","name":"Hidden Functionality"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01","tags":[]},{"url":"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication","tags":[]},{"url":"https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor","tags":[]},{"url":"https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/","tags":[]}],"techniques":[],"reports":[],"affected":[],"advisories":[],"euvd":{"id":"EUVD-2025-1797","url":"https://euvd.enisa.europa.eu/vulnerability/EUVD-2025-1797","exploited":false,"exploitedSince":null,"provenance":"source"},"exploits":[],"exploitSources":null,"exploitRefs":[],"flow":[],"timeline":[{"date":"2025-01-30","kind":"published","label":"Published (CVE.org)","url":"https://www.cve.org/CVERecord?id=CVE-2025-0626","provenance":"source"},{"date":"2026-01-20","kind":"exploited-reported","label":"Exploitation first reported (VulnCheck KEV evidence)","url":"https://www.loginsoft.com/reports/annually/vulnerability-intelligence-report-2025","provenance":"source"}],"speed":{"daysToPublicExploit":null,"firstExploitFrom":null,"daysToKev":null},"priority":{"score":59.8,"factors":[{"name":"In a KEV catalog","points":35.0,"detail":"listed in VulnCheck KEV"},{"name":"Recent KEV listing","points":5.0,"detail":"added 305 days ago"},{"name":"Known ransomware use","points":0.0,"detail":"not known"},{"name":"EPSS percentile","points":13.49,"detail":"percentile 0.6743"},{"name":"Named in threat reports","points":0.0,"detail":"0 reports"},{"name":"Public exploit code","points":0.0,"detail":"none found"},{"name":"Mentioned in OTX pulses","points":1.08,"detail":"2 pulses"},{"name":"CVSS severity","points":5.25,"detail":"base score 7.5"},{"name":"End-of-life product","points":0.0,"detail":"supported or unknown"}]}}