{"id":"CVE-2025-0683","published":"2025-01-30T18:17:35Z","modified":"2025-02-12T20:41:36Z","title":"Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor","vendor":"Contec Health","product":"CMS8000 Patient Monitor","tier":"core","eol":null,"affectedVersions":[{"vendor":"Contec Health","product":"CMS8000 Patient Monitor","versions":["All versions"]}],"solutions":{"items":["Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."],"from":"CVE.org CNA","provenance":"source"},"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"partial","timestamp":"2025-01-30T19:06:59.525559Z","from":"CISA ADP","provenance":"source"},"description":{"text":"In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario.","provenance":"source","from":"CVE.org"},"summary":null,"cvss":{"score":5.9,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00775,"percentile":0.56688,"date":"2026-11-20","provenance":"source"},"kev":null,"vulncheckKev":{"dateAdded":"2026-01-20","exploitedFirstSeen":"2026-01-20","ransomware":false,"evidence":[{"url":"https://www.loginsoft.com/reports/annually/vulnerability-intelligence-report-2025","date":"2026-01-20"},{"url":"https://www.trellix.com/assets/reports/trellix-healthcare-cybersecurity-threat-intelligence-report.pdf","date":"2026-01-27"}],"evidenceTotal":2,"provenance":"source"},"exploited":true,"otx":{"pulseCount":3,"pulses":[{"name":"Healthcare Sector Threat Intel - FDA 510(k) Medical Device Security (DugganUSA)","url":"https://otx.alienvault.com/pulse/692c9d81b4e8f82f6d98e1c6","created":"2025-11-30"},{"name":"Contec CMS8000 Contains a Backdoor","url":"https://otx.alienvault.com/pulse/67a190facf12f98043d32b8a","created":"2025-02-04"},{"name":"Contec CMS8000 Contains a Backdoor","url":"https://otx.alienvault.com/pulse/67a06a13cefb52d72853cb73","created":"2025-02-03"}],"provenance":"machine"},"cwes":[{"id":"CWE-359","name":"Exposure of Private Personal Information to an Unauthorized Actor"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01","tags":[]},{"url":"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication","tags":[]},{"url":"https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor","tags":[]},{"url":"https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/","tags":[]}],"techniques":[],"reports":[],"affected":[],"advisories":[],"euvd":{"id":"EUVD-2025-1815","url":"https://euvd.enisa.europa.eu/vulnerability/EUVD-2025-1815","exploited":false,"exploitedSince":null,"provenance":"source"},"exploits":[],"exploitSources":null,"exploitRefs":[],"flow":[],"timeline":[{"date":"2025-01-30","kind":"published","label":"Published (CVE.org)","url":"https://www.cve.org/CVERecord?id=CVE-2025-0683","provenance":"source"},{"date":"2026-01-20","kind":"exploited-reported","label":"Exploitation first reported (VulnCheck KEV evidence)","url":"https://www.loginsoft.com/reports/annually/vulnerability-intelligence-report-2025","provenance":"source"}],"speed":{"daysToPublicExploit":null,"firstExploitFrom":null,"daysToKev":null},"priority":{"score":56.8,"factors":[{"name":"In a KEV catalog","points":35.0,"detail":"listed in VulnCheck KEV"},{"name":"Recent KEV listing","points":5.0,"detail":"added 305 days ago"},{"name":"Known ransomware use","points":0.0,"detail":"not known"},{"name":"EPSS percentile","points":11.34,"detail":"percentile 0.5669"},{"name":"Named in threat reports","points":0.0,"detail":"0 reports"},{"name":"Public exploit code","points":0.0,"detail":"none found"},{"name":"Mentioned in OTX pulses","points":1.37,"detail":"3 pulses"},{"name":"CVSS severity","points":4.13,"detail":"base score 5.9"},{"name":"End-of-life product","points":0.0,"detail":"supported or unknown"}]}}