{"id":"CVE-2025-14096","published":"2025-12-17T12:19:27Z","modified":"2025-12-17T14:43:16Z","title":"Credential Disclosure vulnerability in Radiometer Products","vendor":"Radiometer Medical Aps","product":"ABL90 FLEX and ABL90 FLEX PLUS Analyzers","tier":"tail","description":{"text":"A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE's are CVE-2025-14095 & CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.","provenance":"source","from":"CVE.org"},"cvss":{"score":8.4,"severity":"HIGH","version":"3.1","vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00222,"percentile":0.07543,"provenance":"source"},"cwes":[{"id":"CWE-798","name":"Use of Hard-coded Credentials"},{"id":"CWE-250","name":"Execution with Unnecessary Privileges"}],"references":[{"url":"https://www.radiometer.com/myradiometer","tags":[]}],"priority":{"score":7.4,"factors":[{"name":"EPSS percentile","points":1.51},{"name":"CVSS severity","points":5.88}]},"vendorSlug":"radiometer-medical-aps","affectedVersions":[{"vendor":"Radiometer Medical Aps","product":"ABL90 FLEX and ABL90 FLEX PLUS Analyzers","versions":["Windows 7 Operating system","Windows XP Operating system"]},{"vendor":"Radiometer Medical Aps","product":"AQT90 FLEX Analyzers","versions":["Windows 7 Operating system","Windows XP Operating system"]},{"vendor":"Radiometer Medical Aps","product":"ABL800 BASIC and ABL800 FLEX analyzers","versions":["Windows 7 Operating system","Windows XP Operating system"]}],"solutions":{"items":["Local Radiometer representatives will contact all affected customers to discuss a permanent solution."],"from":"CVE.org CNA","provenance":"source"},"workarounds":{"items":["Only authorized people can physically access the analyzer."],"from":"CVE.org CNA","provenance":"source"},"capec":[{"id":"CAPEC-69","name":"— Target Programs with Elevated Privileges"}],"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total","timestamp":"2025-12-17T14:42:57.843796Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2025-203889","exploited":false,"provenance":"source"}}