{"id":"CVE-2025-24836","published":"2025-02-13T21:55:12Z","modified":"2025-02-14T18:11:19Z","title":"Qardio Heart Health IOS and Android Application and QardioARM A100 Uncaught Exception","vendor":"Qardio","product":"Heart Health IOS Mobile Application","tier":"tail","description":{"text":"With a specially crafted Python script, an attacker could send \ncontinuous startMeasurement commands over an unencrypted Bluetooth \nconnection to the affected device. This would prevent the device from \nconnecting to a clinician's app to take patient readings and ostensibly \nflood it with requests, resulting in a denial-of-service condition.","provenance":"source","from":"CVE.org"},"cvss":{"score":7.1,"severity":"HIGH","version":"3.1","vector":"CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00253,"percentile":0.17103,"provenance":"source"},"cwes":[{"id":"CWE-248","name":"Uncaught Exception"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-044-01","tags":[]},{"url":"https://www.qardio.com/about-us/#contact","tags":[]}],"priority":{"score":8.4,"factors":[{"name":"EPSS percentile","points":3.42},{"name":"CVSS severity","points":4.97}]},"vendorSlug":"qardio","affectedVersions":[{"vendor":"Qardio","product":"Heart Health IOS Mobile Application","versions":["2.7.4"]},{"vendor":"Qardio","product":"Heart Health Android Mobile Application","versions":["2.5.1"]},{"vendor":"Qardio","product":"QardioARM","versions":["All versions"]}],"workarounds":{"items":["Qardio has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Qardio customer support https://www.qardio.com/about-us/#contact for additional information. Users should do the following to help mitigate the risk: * Disable Bluetooth when not in use. * Don't use this device in public or within Bluetooth range of malicious actors. * Only use trusted mobile apps from trusted providers."],"from":"CVE.org CNA","provenance":"source"},"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total","timestamp":"2025-02-14T18:11:06.923716Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2025-3959","exploited":false,"provenance":"source"}}