{"id":"CVE-2025-4393","published":"2025-07-24T03:22:20Z","modified":"2026-03-27T19:38:42Z","title":"Medtronic MyCareLink Patient Monitor Deserialization Vulnerability","vendor":"Medtronic","product":"MyCareLink Patient Monitor 24950","tier":"tail","description":{"text":"Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to crash the service or elevate privileges. \n\nThis issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025","provenance":"source","from":"CVE.org"},"cvss":{"score":6.5,"severity":"MEDIUM","version":"3.1","vector":"CVSS:3.1/AV:L/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00194,"percentile":0.09251,"provenance":"source"},"cwes":[{"id":"CWE-502","name":"Deserialization of Untrusted Data"}],"references":[{"url":"https://www.medtronic.com/en-us/e/product-security/security-bulletins/mycarelink-patient-monitor-vulnerabilities.html","tags":["vendor-advisory"]},{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-205-01","tags":["third-party-advisory"]}],"priority":{"score":6.4,"factors":[{"name":"EPSS percentile","points":1.85},{"name":"CVSS severity","points":4.55}]},"vendorSlug":"medtronic","affectedVersions":[{"vendor":"Medtronic","product":"MyCareLink Patient Monitor 24950","versions":["< June 25, 2025"]},{"vendor":"Medtronic","product":"MyCareLink Patient Monitor 24952","versions":["< June 25, 2025"]}],"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total","timestamp":"2025-07-24T14:33:13.440835Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2025-22528","exploited":false,"provenance":"source"}}