{"id":"CVE-2025-64298","published":"2025-12-02T21:05:38Z","modified":"2025-12-09T17:03:09Z","title":"Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource","vendor":"Mirion Medical","product":"EC2 Software NMIS BioDose","tier":"tail","description":{"text":"NMIS/BioDose V22.02 and previous version installations where the embedded Microsoft SQLServer Express is used are exposed in the Windows share accessed by clients in networked installs. By default, this directory has insecure directory paths that allow access to the SQL Server database and configuration files, which can contain sensitive data.","provenance":"source","from":"CVE.org"},"cvss":{"score":8.4,"severity":"HIGH","version":"3.1","vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00461,"percentile":0.31983,"provenance":"source"},"cwes":[{"id":"CWE-732","name":"Incorrect Permission Assignment for Critical Resource"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-336-01","tags":[]}],"priority":{"score":12.3,"factors":[{"name":"EPSS percentile","points":6.4},{"name":"CVSS severity","points":5.88}]},"vendorSlug":"mirion-medical","affectedVersions":[{"vendor":"Mirion Medical","product":"EC2 Software NMIS BioDose","versions":["< 23.0"]}],"fixedVersions":[{"vendor":"Mirion Medical","product":"EC2 Software NMIS BioDose","versions":["23.0"]}],"solutions":{"items":["Mirion Medical recommends users update to V23.0 or later. Users with an active support contract should update to the latest version through the software or users can contact Mirion Medical support directly."],"from":"CVE.org CNA","provenance":"source"},"ssvc":{"exploitation":"none","automatable":"no","technicalImpact":"total","timestamp":"2025-12-09T17:03:03.295083Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2025-200323","exploited":false,"provenance":"source"}}