{"id":"CVE-2026-105279","published":"2026-10-16T14:30:51Z","modified":"2026-10-16T15:20:18Z","title":"J4Care dcm4che Missing Authentication for Critical Function","vendor":"J4Care","product":"dcm4chee-arc-light","tier":"tail","description":{"text":"In the default configuration, the archive accepts DICOM associations from an unregistered calling AE. A remote peer can use C-FIND and C-GET to enumerate and retrieve patient information and medical images. It can also use C-STORE to file fabricated study content under an existing patient identity.","provenance":"source","from":"CVE.org"},"cvss":{"score":9.1,"severity":"CRITICAL","version":"3.1","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.0042,"percentile":0.28053,"provenance":"source"},"cwes":[{"id":"CWE-306","name":"Missing Authentication for Critical Functionality"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-288-02","tags":[]},{"url":"https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-288-02.json","tags":[]}],"priority":{"score":12.0,"factors":[{"name":"EPSS percentile","points":5.61},{"name":"CVSS severity","points":6.37}]},"vendorSlug":"j4care","affectedVersions":[{"vendor":"J4Care","product":"dcm4chee-arc-light","versions":["5.35.1"]}],"fixedVersions":[{"vendor":"J4Care","product":"dcm4chee-arc-light","versions":["5.35.2"]}],"solutions":{"items":["J4Care has provided a fix for the vulnerability. J4Care recommends users implement release 5.35.2 or later."],"from":"CVE.org CNA","provenance":"source"},"ssvc":{"exploitation":"none","automatable":"yes","technicalImpact":"total","timestamp":"2026-10-16T15:05:44.341719Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2026-102894","exploited":false,"provenance":"source"}}