{"id":"CVE-2026-78224","published":"2026-09-11T14:20:06Z","modified":"2026-09-11T14:57:02Z","title":"NextGen Healthcare Mirth Connect Improper Restriction of XML External Entity Reference","vendor":"NextGen Healthcare","product":"Mirth Connect","tier":"tail","description":{"text":"The XSLT Transformer Step builds a bare TransformerFactory without the proper security options set, so XXE injection can allow data exfiltration and denial-of-service attacks.","provenance":"source","from":"CVE.org"},"cvss":{"score":8.2,"severity":"HIGH","version":"3.1","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:L","provenance":"source","from":"CVE.org CNA"},"epss":{"score":0.00437,"percentile":0.29906,"provenance":"source"},"cwes":[{"id":"CWE-611","name":"Improper Restriction of XML External Entity Reference"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-253-01","tags":[]},{"url":"https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-253-01.json","tags":[]}],"priority":{"score":11.7,"factors":[{"name":"EPSS percentile","points":5.98},{"name":"CVSS severity","points":5.74}]},"vendorSlug":"nextgen-healthcare","affectedVersions":[{"vendor":"NextGen Healthcare","product":"Mirth Connect","versions":["<= 4.7.1"]}],"fixedVersions":[{"vendor":"NextGen Healthcare","product":"Mirth Connect","versions":["4.7.2"]}],"solutions":{"items":["NextGen recommends users update Mirth Connect v4.7.2 or later. Users can download the latest version from the NextGen Healthcare customer portal."],"from":"CVE.org CNA","provenance":"source"},"ssvc":{"exploitation":"none","automatable":"yes","technicalImpact":"partial","timestamp":"2026-09-11T14:51:00.487459Z","from":"CISA ADP","provenance":"source"},"euvd":{"id":"EUVD-2026-76116","exploited":false,"provenance":"source"}}