Patchfirst
CVE-2024-12248 Critical · 9.8 EPSS 1.3%

Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor

Fix:see the vendor's fix instructions Instructions ↓

Contec Health · CMS8000 Patient Monitor

Summary

The out-of-bounds write flaw lives in the Contec Health CMS8000 Patient Monitor, a patient monitoring device. An attacker who can reach the device over the network can send specially formatted UDP requests that write arbitrary data, potentially achieving remote code execution, with no authentication or user interaction required and no special configuration mentioned.

Official description · CVE.org

Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.

Breakdown
61.7 / 100
Exploited
Yes, in VulnCheck KEV
Reports
0
Tactics
0/15

First seen exploited 12 months after disclosure.

  1. Disclosedday 0
  2. First seen exploitedVulnCheck KEV evidence+12 mo

Affected versions and fixes

as published in the CVE record

Contec Health CMS8000 Patient Monitor

Affected: Firmware version smart3250-2.6.27-wlan2.1.7.cramfs · Firmware version CMS7.820.075.08/0.74(0.75) · Firmware version CMS7.820.120.01/0.93(0.95)

Fix

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .

What security researchers reported

Not named in any threat report yet