CVE-2025-0626 High · 7.5 EPSS 1.1%

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

Workaround:the vendor lists a workaround. No fixed version in the CVE record. See workaround ↓

Contec Health · CMS8000 Patient Monitor

Official description · CVE.org

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Breakdown
59.8 / 100
Exploited
Yes, in VulnCheck KEV
Reports
0
Tactics
0/15

First seen exploited 12 months after disclosure.

  1. Disclosedday 0
  2. First seen exploitedVulnCheck KEV evidence+12 mo

Affected versions and fixes

as published in the CVE record

Contec Health CMS8000 Patient Monitor

Affected: All versions

Workaround

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .