Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Workaround:the vendor lists a workaround. No fixed version in the CVE record. See workaround ↓
Contec Health · CMS8000 Patient Monitor
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
- Exploited
- Yes, in VulnCheck KEV
- Reports
- 0
- Tactics
- 0/15
Exploitation timelineDated milestones for a CVE: when it was disclosed, when exploit code became public, when exploitation was confirmed, and when threat reports named it.
First seen exploited 12 months after disclosure.
- Disclosedday 0
- First seen exploitedVulnCheck KEV evidence+12 mo
Affected versions and fixes
as published in the CVE recordContec Health CMS8000 Patient Monitor
Affected: All versions
Workaround
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .
What security researchers reported
Not named in any threat report yetAlso mentioned in 2 community threat pulses (AlienVault OTX)
Weaker evidencePulses are indicator collections that anyone can publish. A mention is not confirmation of exploitation.
- Contec CMS8000 Contains a Backdoor4 Feb 2025
- Contec CMS8000 Contains a Backdoor3 Feb 2025