CVE-2025-0683 Medium · 5.9 EPSS 0.8%

Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

Fix:see the vendor's fix instructions Instructions ↓

Contec Health · CMS8000 Patient Monitor

Official description · CVE.org

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Breakdown
56.8 / 100
Exploited
Yes, in VulnCheck KEV
Reports
0
Tactics
0/15

First seen exploited 12 months after disclosure.

  1. Disclosedday 0
  2. First seen exploitedVulnCheck KEV evidence+12 mo

Affected versions and fixes

as published in the CVE record

Contec Health CMS8000 Patient Monitor

Affected: All versions

Fix

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .