Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
Fix:see the vendor's fix instructions Instructions ↓
Contec Health · CMS8000 Patient Monitor
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
- Exploited
- Yes, in VulnCheck KEV
- Reports
- 0
- Tactics
- 0/15
Exploitation timelineDated milestones for a CVE: when it was disclosed, when exploit code became public, when exploitation was confirmed, and when threat reports named it.
First seen exploited 12 months after disclosure.
- Disclosedday 0
- First seen exploitedVulnCheck KEV evidence+12 mo
Affected versions and fixes
as published in the CVE recordContec Health CMS8000 Patient Monitor
Affected: All versions
Fix
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .